ClearanceBrief
Recall Z-1684-2024 ยท DQO Catheter, Intravascular, Diagnostic

Boston Scientific Corporation

Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AL2 (5PK), REF H74908526982; cardiac catheter

DateMay 10, 2024 (initiated Mar 21, 2024)
StatusOpen, Classified
ReasonAn increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
Root causeUnder Investigation by firm
ActionBoston Scientific issued an Urgent Medical Device Removal notice to its consignees on 03/21/2024 via overnight delivery. The notice explained the issue, the potential risk, and requested the removal of the affected products. Distributors were directed to notify their consignees.
Quantity8715 catheters
DistributionWorldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152605Impulse Angiographic Catheter, Expo Angiographic Catheter
Boston Scientific Corporation, Oct 29, 2015