ClearanceBrief
Recall Z-2420-2024 · MEA Pump, Infusion, Pca

Smiths Medical ASD Inc.

CADD-Solis VIP ambulatory Infusion Pumps, Model 2110. Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that require a continuous rate of infusion, intermittent bolus, and/or with patient-controlled demand doses.

DateAug 3, 2024 (initiated Feb 27, 2024)
StatusOpen, Classified
ReasonThe CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Root causeSoftware Manufacturing/Software Deployment
ActionSmiths medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/27/2024 via letter distributed using traceable means. The notice explained the issues affecting the devices, potential risks, and requested the following: "Required Actions for Users: 1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations. 2. Complete and return the attached Response Form to smithsmedical3563@sedgwick.com within ten days of receipt to acknowledge your understanding
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110; PharmGuard® Ad
Smiths Medical Ads, Inc., Aug 24, 2017