Recall Z-1507-2024 ยท JAK System, X-Ray, Tomography, Computed

Philips North America

PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.1.41989-Intended to produce cross-sectional images of the body Model Number: 728143

DateApr 9, 2024 (initiated Dec 24, 2020)
StatusOpen, Classified
ReasonPatient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured
Root causeSoftware design
ActionPhilips Healthcare issued Customer Information letter on 12/24/20 via USPS. Letter states reason for recall, health risk and action to take: You may continue to use your system(s) in accordance with the intended use. Refer to ADVICE ACTIONS BY CUSTOMER/USER of the customer letter for specific details regarding the issue descriptions and advice to customers. Should the customer feel uncertain regarding this action, please contact your local Philips representative. Field Change Orders (FCO72800727) involving software upgrade will be released to correct the issue at no charge to you. The planned
Quantity428 units
DistributionNationwide including Puerto Rico Foreign: Albania Argentina Australia Austria Bangladesh Bolivia Brazil Bulgaria China Czech Republic Denmark Ecuador Egypt France Germany Greece Hungary India Indonesia Iraq Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Libya Malaysia M
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180015Philips Incisive CT
Philips Healthcare (Suzhou) Co., Ltd., Mar 20, 2018