Recall Z-1329-2024 ยท HST Apparatus, Traction, Non-Powered

Baxter Healthcare Corporation

HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery

DateMar 20, 2024 (initiated Mar 11, 2024)
StatusOpen, Classified
ReasonBaxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
Root causeUnder Investigation by firm
ActionBaxter issued an Urgent Medical Device Recall notice to its sole consignee on 03/11/2024 via email. The notice explained the issue and potential hazard and requested the following: 1. Immediately locate, isolate, and cease all use of the affected serial numbers of the product. The product code and serial number can be found on labelling on the bottom of the boot. 2. Arrange for the return of your impacted product by calling the Baxter Order Management Team at 800-433-5774, option 2 then 3, between the hours of 8:00 am and 5:00 pm Eastern Time, Monday through Friday, or by emailing Baxter at
Quantity80 units
DistributionUS Nationwide distribution in the state of Florida.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.