ClearanceBrief
Recall Z-1506-2024 ยท FSY Light, Surgical, Ceiling Mounted

Stryker Communications

Chromophare Surgical Light System , REF CH00000001

DateApr 9, 2024 (initiated Feb 8, 2024)
StatusOpen, Classified
ReasonDue to insufficient mounting force the stability of surgical light system cannot be guaranteed.
Root causeEmployee error
ActionOn February 19, 2024, Stryker issued a Urgent Medical Device Recall notification via XXX. In addition to notifying consignees about the recall. Stryker asked consignees to take the following actions: Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the product listed in the product affected grid. If affected product is found, segregate and do not use the product. 3. Complete and sign the enclosed Business Reply Form on Attachment A and return to CommunicationsRA1@stryker.com. 4. A Sales Representative or Field Service
Quantity1 unit
DistributionUS: None OUS: Chili
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120392CHROMOPHARE F 628 AND F 528
Berchtold GmbH & Co. KG, Feb 16, 2012