Recall Z-1398-2024 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Becton Dickinson & Co.

BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with car

DateMar 29, 2024 (initiated Feb 29, 2024)
StatusOpen, Classified
ReasonProducts were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results
Root causeUnder Investigation by firm
ActionBD issued URGENT: Medical Device Product Correction Product Removal letter on 2/29/24 via courier service via tracked letter and/or e-mail. Letter states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood. 3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the
Quantity11,800 eaches
DistributionWorldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132674BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- MEROPENEM (0.125-32 UG/ML)
Becton, Dickinson and Company, Dec 27, 2013