Recall Z-1392-2024 ยท QRY Dental Navigation System

X-NAV Technologies, LLC

X-Guide Handpiece Adaptor Sleeve 1, Model P010701

DateMar 29, 2024 (initiated Mar 4, 2024)
StatusOpen, Classified
ReasonThe Handpiece Adaptor may have a manufacturing defect causing incorrect geometry. The incorrect geometry in turn causes the Handpiece Adaptor to not fit onto the doctor's dental handpiece.
Root causeUse error
ActionAn URGENT MEDICAL DEVICE RECALL dated 3/4/24 was sent to customers. Action To Be Taken by the User Identify if you possess any of the affected devices. The lot numbers of the affected devices are the following: 2308010012 2306020012 2310010012 If the lot number on your device is not one of the above, then that Adaptor Sleeve is not part of this Recall. If you have a defective device, quarantine the device so that it is prevented from being used. Finally, destroy the device or return to X-Nav Technologies, LLC. Return the response form. X-Nav will ship all affected users a new Handpiece Ada
Quantity88 units
DistributionWorldwide distribution - US Nationwide and the countries of China, Europe, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232148X-Guide Surgical Navigation System
X-Nav Technologies, LLC, Feb 21, 2024