ClearanceBrief
Recall Z-1351-2024 ยท BZD Ventilator, Non-Continuous (Respirator)

SleepNet Corporation

Mojo Full Face with Headgear, All Sizes

DateApr 1, 2024 (initiated Mar 1, 2024)
StatusOpen, Classified
ReasonUpdate to contraindications and warning language due to CPAP masks containing magnets.
Root causeDevice Design
ActionThe firm emailed consignees a notification titled URGENT: MEDICAL DEVICE RECALL, dated 3/1/24. Consignees were instructed to stop using recalled masks if the user or anyone in close physical contact with the mask has an active medical implant or metallic implant that will interact with magnets and contact their mask provider for a replacement mask that does not contain magnets. Consignees that do not have an active medical implant or metallic implant that will interact with magnets do not have to cease use. If a consignee is unsure whether or not to cease use, they should consult with their ph
Quantity23,042 units
DistributionWorldwide distribution - US Nationwide and the countries of AUSTRIA, BAHRAIN, BRAZIL, COLOMBIA, COSTA RICA, ECUADOR, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, INDIA, IRAQ, IRELAND, ISRAEL, ISREAL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NORWAY, POLAND, Q
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120463MOJO, VERASEAL
Sleepnet Corporation, Jul 27, 2012