ClearanceBrief
Recall Z-1515-2024 ยท KNT Tubes, Gastrointestinal (And Accessories)

Boston Scientific Corporation

EndoVive 20Fr Push Safety PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00566471

DateApr 11, 2024 (initiated Feb 20, 2024)
StatusOpen, Classified
ReasonInability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration
Root causeProcess design
ActionBoston Scientific issued Urgent Medical Device Removal letter on 2/20/24 via courier. Letter states reason for recall, health risk and action to take: is readily accessible to all handlers and users of the device. Share this notice with any health care professional within your organization and with any organization where the affected devices may have been transferred. This removal affects only the products listed in Attachment 1 (Affected Products). No other material numbers or batches are impacted by this removal. Further distribution or use of any remaining product affected by this removal s
Quantity2501 units
DistributionWorldwide distribution - US Nationwide and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031538ENDOVIVE INITIAL PLACEMENT PEG KIT, DIRECT PEJ KIT, AND PEG SAFETY KIT
Boston Scientific Corp, Jun 27, 2003