Recall Z-1605-2024 ยท NKB Thoracolumbosacral Pedicle Screw System

SEASPINE ORTHOPEDICS CORPORATION

Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXT

DateApr 22, 2024 (initiated Feb 15, 2024)
StatusOpen, Classified
ReasonScrew heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Root causeNonconforming Material/Component
ActionOn 2/14/24 the firm phoned distributors and informed them of the recall issue and next steps. On 2/16/24 the firm sent a recall communication via email. Customers were asked to arrange for the return of these sets along with any additional Mariner RDX inventory using a firm provided return label. Customers with questions or concerns can contact the Associate Product Manager, at olivia.disanto@seaspine.com
Quantity1579 Screw Heads
DistributionUS Nationwide distribution in the states of MI, GA, VA, IN, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222110Mariner RDX System
SeaSpine Orthopedics Corporation, Sep 7, 2022