Recall Z-1482-2024 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.

DateApr 5, 2024 (initiated Feb 23, 2024)
StatusOpen, Classified
ReasonA use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
Root causeUnder Investigation by firm
ActionOn February 26, 2024, the firm notified customers of the issue via Urgent Field Safety Notice: Medical Device Correction letters via email. Actions to be taken by the user: 1. Educate planning staff and all users about the possibility of the described use error. 2. If the use error is deemed likely to occur at your clinic, take actions to ensure that proper review of assigned materials is performed before a patient model is used for treatment planning. 3. If you wish to have the Silicon chemical element removed from the material list, contact RaySearch support for assistance. This is especiall
Quantity154 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190387RayStation
RaySearch Laboratories AB (PUBL), Jun 19, 2019
K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020
K210645RayStation 10.1
RaySearch Laboratories AB (PUBL), Jun 29, 2021
K211867RayStation 11.0
RaySearch Laboratories AB (PUBL), Sep 8, 2021
K220141RayStation 11B
RaySearch Laboratories AB (PUBL), Apr 15, 2022
K222312RayStation 12A
RaySearch Laboratories AB (PUBL), Mar 29, 2023