Recall Z-1435-2024 ยท GDG Hook, Surgical, General & Plastic Surgery
MEDLINE INDUSTRIES, LP - Northfield
Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F), Product Code 66660; b) STERILE SMALL 2 PRONG SKIN HOOK (25754), Product Code 67105; c) STERILE NEWS TRACH HOOK(83300), Product Code 67810; d) STERILE LILLIE EAR HOOK DELICATE 7", Product Code I68655
| Date | Mar 29, 2024 (initiated Feb 23, 2024) |
| Status | Open, Classified |
| Reason | Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users. |
| Root cause | Process control |
| Action | Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers. |
| Quantity | 1520 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Panama, Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.