Recall Z-1486-2024 ยท FSA Lift, Patient, Non-Ac-Powered

Human Care Hc Sweden Ab

Altair 330 lbs w/o Hanger Bar, REF: 55200H-NA, 55220H-NA, and Roomer S 220 kg w/o Hanger Bar, REF: 55300H-NA, patient lifts

DateApr 5, 2024 (initiated Feb 23, 2024)
StatusOpen, Classified
ReasonPatient lift hanger bar may come loose, which could potentially pose risks to patients.
Root causeProcess control
ActionOn 2/23/24, Field Safety Notices were emailed to customers who were asked to do the following: 1) Checking the fixation by carefully pulling on the shaft. Should you observe any deviation or irregularity, please inform us promptly. 2) Read and understand this letter and provide a confirmation via email to the address quality@humancaregroup.com In case of recognizing any deviation or irregularity during the fixation control or if in need of additional information email quality@humancaregroup.com On 5/6/24, Field Safety Notices were emailed to customers with additional instructions for distr
Quantity10
DistributionUS Nationwide distribution in the states of TX, IL, OK, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.