ClearanceBrief
Recall Z-1594-2024 ยท DDZ Albumin, Fitc, Antigen, Antiserum, Control

Randox Laboratories Ltd.

Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

DateApr 17, 2024 (initiated Jan 31, 2024)
StatusOpen, Classified
ReasonMicroalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Root causeUnder Investigation by firm
ActionRandox Laboratories (Manufacturer) issued Urgent Field Safety Notice on 31 Jan 24 Complaint Reference: REC716 letter via email to Distributors. Distributors will contact the end-users. Letter states reason for recall, health risk and action to take: We have reassigned the target values in the above lot numbers in line with the ERM reference material. Please refer to the table below for the updated calibrator targets. You may experience a shift in Quality Control and patient sample recovery of up to +12%. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the u
Quantity6 units
DistributionUS Nationwide distribution in the states of GA, MA, WV including PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002674RANDOX MICROALBUMIN
Randox Laboratories, Ltd., Oct 5, 2000