Recall Z-1330-2024 ยท GDY Gauze/Sponge, Internal, X-Ray Detectable

Covidien

Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1

DateMar 21, 2024 (initiated Feb 21, 2024)
StatusOpen, Classified
ReasonPacks including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
Root causeLabeling mix-ups
ActionOn 2/21/24, recall notices were mailed to customers and distributors who were asked to do the following: 1) Do not use impacted products. 2) Remove, quarantine, and return all unused affected product. 3) This communication needs to be passed to all those who need to be aware within your organization where impacted devices have been transferred, and to any other associated organizations that may be impacted by this action. 4) Complete and return the customer confirmation form to rs.gmbmitgfca@medtronic.com Customers with questions can contact the firm at 800-962-9888 Option 2.
Quantity755,800
DistributionUS Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.