Recall Z-1307-2024 ยท LNZ Changer, Tube, Endotracheal

Sarnova HC, Llc

Curaplex Nasal Intubation Kit with 6mm Endotrol Tube-To perform a nasal intubation to establish an airway in an emergency. Item Number: 023060S

DateMar 19, 2024 (initiated Feb 12, 2024)
StatusOpen, Classified
ReasonEndotracheal Tube Holder included in certain kits is not compatible with a nasal intubation procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
Root causeComponent design/selection
ActionBound Tree issued URGENT NOTICE OF RECALL letters on February 12, 2024 via USPS Letter to Medical Inventory Controls/Supplies Coordinator/Buyer. Letter states reason for recall, health risk and actin to take: Should you determine your inventory contains product covered by this recall, please immediately quarantine your affected product and contact Bound Tree Medical Customer Care at 800.533.0523 to arrange for a return of the product in exchange for a credit
Quantity540 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.