ClearanceBrief
Recall Z-1269-2024 ยท LLZ System, Image Processing, Radiological

FUJIFILM Healthcare Americas Corporation

Synapse PACS - Version 7.1.000US

DateMar 7, 2024 (initiated Jan 10, 2024)
StatusOpen, Classified
ReasonMeasurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
Root causeSoftware design
ActionURGENT: MEDICAL DEVICE CORRECTION letters were sent to customers January 26, 2024. In an abundance of caution, FUJIFILM is recommending that customer/end users take one of the following actions until the correction is applied: 1. To not perform measurements on Secondary Capture Images that are displayed together with Breast Tomo Images. 2. Ensure Secondary Capture Image has valid pixel spacing values using the DICOM Header Tool. Please complete and return the Field Action Verification Form attached on page 3 of this communication. Providing the response with the information requested is essent
Quantity3 Units (3 US)
DistributionWorldwide distribution - US Nationwide, including Puerto Rico and the countries of Australia, Belgium, Brazil, Canada, Colombia, Ecuador, France, Germany, India, Italy, Mexico, Namibia, Netherlands, Portugal, Qatar, Spain, Thailand, Turkey, UAE, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190232Synapse PACS
Fujifilm Corporation, Mar 26, 2019