Recall Z-1275-2024 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Smiths Medical ASD Inc.

smiths medical Pneupac paraPAC plus 300 ventilator kit without PEEP and CPAP, REF P300NUS

DateMar 19, 2024 (initiated Feb 7, 2024)
StatusOpen, Classified
ReasonWhen a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous positive gas flow when in the cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.
Root causeUnder Investigation by firm
ActionSmiths medical issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 02/07/2024 via letter using traceable means. The notice explained the issue with the device, potential risk to the patient, and requested the following actions be taken: "There is no need to return or discontinue using your paraPAC plus, at this time. When using the device, all instructions, including warnings and cautions in the User Manual Doc. numbers (10018833-003 and/or 10026347-002) must be followed with heightened awareness. This is inclusive, but not limited to the following: " Constant monitorin
Quantity2902 units
DistributionWorldwide distribution: US (nationwide) and OUS countries of: Algeria, Armenia, Australia, Austria, Bangladesh, Bahrain, Bolivia, Plurinational state of, Brazil, Bulgaria, Canada, Colombia, Czech Republic, Cyprus, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hong Kong, Iceland, Ind
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123957PNEUPAC PARAPAC PLUS
Smiths Medical International , Ltd., Sep 16, 2013