ClearanceBrief
Recall Z-1326-2024 · KDJ Set, Administration, For Peritoneal Dialysis, Disposable

Fresenius Medical Care Holdings, Inc.

stay-safe /Luer Lock Catheter Ext. 18 in

DateMar 26, 2024 (initiated Jan 23, 2024)
StatusOpen, Classified
ReasonThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Root causeDevice Design
ActionIMPORTANT MEDICAL DEVICE CORRECTION Letter were sent to customer on February 14, 2024 informing customers that non-dioxin-like polychlorinated biphenyl acids (PCBAs) could leach from the stay-safe Catheter Extension sets and Adapter, which are manufactured using peroxide cross-linked silicone tubing. Please note that PCBAs are chemically different from polychlorinated biphenyls (PCBs) which were not identified. No complaints have been reported. PCBA levels decrease with shorter catheter extension sets and decrease over the treatment time. PCBA exposure threshold examples were provided per body
Quantity479,978 eaches
DistributionDomestic: Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022412FRESENIUS CAPD STAY.SAFE DISPOSABLE ADMINISTRATION SETS WITH STAY.SAFE CONNECTOR
Fresenius Medical Care North America, Mar 5, 2003
K173593stay•safe catheter extension set with Safe-Lock, 12 inch, stay•safe catheter extension set with Luer-Lock, 6 inch, stay•
Fresenius Medical Care Renal Therapies Group, LLC, May 21, 2018