ClearanceBrief
Recall Z-1320-2024 ยท HWC Screw, Fixation, Bone

In2Bones, SAS

I.B.S 6.5-C Compression screw - diam 6.5mm lg 80mm,Part Number S65 ST180

DateMar 20, 2024 (initiated Jan 26, 2024)
StatusOpen, Classified
ReasonThis Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
Root causeProcess control
ActionThe firm notified its US consignee via email on 02/26/2024. The email identified the products begin recalled and attached the notice the consignee was to use to notify their customers. That notice, the FIELD CORRECTIVE ACTION, MEDICAL DEVICE BATCH RECALL - EXTENSION, was issued on 01/26/2024, via email, to the US consignees which explained the issue with the device, associated risks, and requested the return of the affected products.
Quantity63 units
DistributionUS Nationwide distribution in the state of Tennessee.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131920STABILIZATION SCREW
In2bones Sas, May 1, 2014