Recall Z-1358-2024 ยท OYX Bcr/Abl1 Monitoring Test

Bio-Rad Laboratories

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

DateMar 22, 2024 (initiated Feb 15, 2024)
StatusOpen, Classified
ReasonDue to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.
Root causeNonconforming Material/Component
ActionOn February 7, 2024, Bio-Rad issued their initial "Urgent Medical Device Correction" notification via E-Mail. On February 14, 2023, Bio RAD Laboratories issued a revised "Urgent Medical Device Correction" notification via E-Mail. Bio-Rad asked consignees to take the following actions: 1. Confirm past patient results against correct values for Catalog No. 12005560, Batch Nos. 64562656 and 64571846 when provided with corrected calibrator values by Bio-Rad Laboratories. 2. If all the controls values fall within the corrected ranges, no further action is required. If one or more controls are o
Quantity33 units
DistributionUS Nationwide distribution in the states of PA, SD, TX, FL, MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181661QXDx BCR-ABL %IS Kit for use on the QXDx AutoDG ddPCR System
Bio-Rad Laboratories, Inc., Feb 13, 2019