Recall Z-1290-2024 ยท DYB Introducer, Catheter

Biosense Webster, Inc.

CARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath is designed to provide accessibility and maneuverability in the cardiac anatomy. The steerable sheath is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with a three-way stopcock is provided for air or blood aspiration, and fluid

DateMar 13, 2024 (initiated Feb 16, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.
Root causeUnder Investigation by firm
ActionOn 02/16/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that Biosense Webster identified through a complaint investigation that a mix-up occurred during the manufacturing of two lots. The devices could potentially be shorter in length (24 1/8 inches (61.28 cm) instead of, 28 inches (71 cm). Customer are instructed/informed: If a CARTO VIZIGO" Bi-Directional Guiding Sheath, Medium Curve (Catalog Number: D-1385-02) from Lot Numbers 60000282 and 60000287 is being used in their facility, they should continue to follow the IFU. The Instruct
Quantity853 units
DistributionU.S. Nationwide distribution including in the states of FL, IL, AL, NY, PA, SC, AZ, MI KY, NJ, WI, IN, LA, TX, MA, OH, AR, MT, MO, VA, NC, TN, CA, MN, WV, CT, CO, IA, MD, ID, NM, NH, UT, GA, WA, NV, OK, KS, AK, NE, HI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170997CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath
Biosense Webster, Inc., Jul 25, 2017