Recall Z-1181-2024 ยท IYE Accelerator, Linear, Medical

PTW-FREIBURG

Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.

DateFeb 22, 2024 (initiated Feb 8, 2024)
StatusOpen, Classified
ReasonSoftware: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
Root causeSoftware design
ActionPTW Dosimetry issued Urgent Medical Device Correction letter via email on 2/8/24. Letter states reason for recall, health risk and action to take: PTW recommends not using the Suppress gamma calculation function for ROIs where a high proportion of voxels are to be suppressed in the gamma calculation. Since the malfunction is limited to the GPR calculation of ROIs and does not affect the GPR calculation of the VERIQA Total volume, it is recommended to rely on the comparison based on VERIQA Total. PTW will correct the malfunction in the next software version. The corrected version will be made
Quantity1 US; 65 worldwide
DistributionGA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213370VERIQA
Ptw-Freiburg Physikalisch-Technische-Werkstaetten, Mar 30, 2022