ClearanceBrief
Recall Z-1306-2024 ยท GYB Media, Electroconductive

Natus Neurology Inc

natus neurology, Sleep Supplies Start Kit with Single Use Supplies for Embla EX Series Amplifiers, REF 024252

DateMar 19, 2024 (initiated Feb 7, 2024)
StatusOpen, Classified
ReasonNatus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five complaints. These customers reported that the gel is too thick, is discolored (darker than usual), and in one instance, patients reported irritation after procedures using the gel.
Root causeUnder Investigation by firm
ActionNatus issued an Urgent Medical Device Recall Notice to its consignees on 02/07/2024 via UPS and email. The notice explained the issue and risk to health. It also requested the consignee cease use, complete and return the form for the replacement of the product.
Quantity35 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K852956CONTACT CONDUCTIVE ADHESIVE GEL COMPOSITION
Parker Laboratories, Inc., Aug 9, 1985