Recall Z-1178-2024 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226

DateFeb 22, 2024 (initiated Feb 6, 2024)
StatusOpen, Classified
ReasonGenerator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
Root causeUnder Investigation by firm
ActionPhilips notified U.S customers by Urgent Medical Device Correction letter on 06-Feb-2024 via certified mailing via United States Postal Service. " Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk, and action to take: Circulate this Letter to all clinical staff and any service personnel that may service the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction Letter with the documentation of the system until Philips implements the correction in your system. " If the Cert
Quantity10 units US; 124 units OUS
DistributionWorldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Repub
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020