Recall Z-1280-2024 ยท GGT Assay, Erythropoietin

Siemens Healthcare Diagnostics, Inc.

Atellica IM Erythropoietin (EPO) Assay (100 Test)

DateMar 11, 2024 (initiated Jan 24, 2024)
StatusOpen, Classified
ReasonSiemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Correction notification letter dated 01/24/24 was sent to customers. Actions to be Taken by the Customer: Discontinue use of and discard the EPO kit lots listed in Table 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued EPO patient testing in your laboratory during this time, please contact your Siemens representative to discuss alternat
Quantity1,057 units (46 US, 1,011 OUS)
DistributionWorldwide - US Nationwide distribution in the states of AL, CA, IA, IL, MI, MN, NC, NE, NV, NY, OH and TX; The countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Italy, Japan, Latvia, Mala
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183088ADVIA Centaur Erythropoietin (EPO) assay
Axis-Shield Diagnostics Limited, Aug 2, 2019