ClearanceBrief
Recall Z-1278-2024 ยท HCI Applier, Aneurysm Clip

Peter Lazic Gmbh

D-Clip Standard/Mini Applier, 210mm bayonet, jaw rigid, double action.

DateMar 11, 2024 (initiated Oct 6, 2023)
StatusOpen, Classified
ReasonCustomer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip.
Root causeDevice Design
ActionReportable Event - Removal letters dated October 6, 2023 were sent to customers. Customers were instructed to return all product subject to recall and to notify any of their customers as well. Please return items for the attention of one of the PRRCs provided to ensure quick processing and assignment to the case. If you have any questions, please contact Mr. Sven Lazic (info@lazic.de +49 (0) 7461 96643 - 20) or Ms. Kim Maier (k.maier@lazic.de or +49 (0) 7461 96643 - 21).
Quantity63 units
DistributionUS Nationwide distribution in the state of ME.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180757D-Clip; D-Clip Applier
Peter Lazic GmbH, Apr 12, 2019