Recall Z-1182-2024 ยท OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

Alphatec Spine, Inc.

IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09

DateFeb 23, 2024 (initiated Jan 26, 2024)
StatusOpen, Classified
ReasonDue to reports of intraoperative graft bolt implantation breakages.
Root causeDevice Design
ActionOn 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this n
Quantity12,777 devices