Recall Z-1202-2024 ยท HBG Drills, Burrs, Trephines & Accessories (Manual)
Integra LifeSciences Corp.
Ref No: INSHITHND / Integra Cranial access kit (without Prep solutions) Bit and Guard. For access to the subarachnoid space or the lateral ventricles of the brain.
| Date | Feb 28, 2024 (initiated Jan 19, 2024) |
| Status | Open, Classified |
| Reason | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Root cause | Process control |
| Action | Integra issued 1st Notification- Urgent: Voluntary Medical Device Recall letter on 01/19/24 via fedex overnight. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service and quarantine. 2. If you do have affected product, check the box on the enclosed Acknowledgement Form I do have affected product. Record the total quantity of the affected product that you have. 3. Forward this notice to those who utilize the device so they are aware of this recall and can identify any affected d |
| Quantity | 915 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Austria, Bahamas, Belgium, Canada, Chile, France, Germany, Greece, Ireland, Israel, Poland, Portugal, Slovenia, South Africa, Spain, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the). |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K961113 | CNS INTERCHANGEABLE BIT CRANIAL DRILL Clinical Neuro Systems, LLC, Jun 5, 1996 |