Recall Z-1595-2024 ยท FNG Lift, Patient, Ac-Powered

Raye's Inc.

SW Bari Lift & Transfer, Model Number 38060000

DateApr 18, 2024 (initiated Jan 25, 2024)
StatusOpen, Classified
ReasonRetrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Root causeDevice Design
ActionSizewise Manufacturing issued an URGENT Voluntary Medical Device Recall to its consignees on 01/29/2024 by letter. The notice explained the issue, potential risk, and requested the discontinuation of use and disposal of the device. The firm sent a follow-up notice on 03/07/2024 asking the distributors to further notify their customers (users of the device).
Quantity1,149 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.