Recall Z-1132-2024 ยท LRP Tray, Surgical
MEDLINE INDUSTRIES, LP - Northfield
Centurion Kits, trays, and packs labeled as follows: a) BASIC PROCEDURE TRAY, REF MNS13600; b) BWH DAILY DRIVELINE MGMT KIT, REF DM640A; c) CIRCUMCISION TRAY, REF CIT6180; d) INCISION & DRAINAGE TRAY, REF ID1690; e) LACERATION TRAY, REF SUT19510; f) LACERATION TRAY, REF SUT19510; g) LACERATION TRAY, REF SUT8565; h) LACERATION TRAY, REF SUT8565; i) LINE REPAIR KIT, REF MNS12045A; j) MMC
| Date | Feb 25, 2024 (initiated Nov 15, 2023) |
| Status | Open, Classified |
| Reason | Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile. |
| Root cause | Nonconforming Material/Component |
| Action | MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different |
| Quantity | 14234 units |
| Distribution | Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.