ClearanceBrief
Recall Z-1172-2024 ยท LLZ System, Image Processing, Radiological

FUJIFILM Healthcare Americas Corporation

Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.

DateFeb 21, 2024 (initiated Jan 3, 2024)
StatusOpen, Classified
ReasonThe LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Root causeSoftware design
ActionOn January 3, 2024, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters. Fujifilm will offer to correct the formula that calculates the LV Mass (2D Bullet) in the Synapse database. Until the update is scheduled and completed, customers may continue to safely use the Pediatric Workflow modules listed in the notice as long as the LV Mass (2D Bullet) equation is not the main factor for a diagnosis. Customers should coordinate with the Fujifilm service engineer to schedule the correction of their device when contacted. Please contact Fujifilm's field action coordinator, Kot
Quantity8
DistributionUS domestic distribution nationwide. International distribution to Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151859Synapse Cardiovascular
Fujifilm Medical Systems U.S.A, Inc., Aug 12, 2015