Recall Z-0978-2024 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Maquet Medical Systems USA

BO-RF-32-USA RotaFlow Centrifugal Pump with SOFTLINE Coating, Product Code 701047553. For use with an extracorporeal cardiovascular or cardiopulmonary bypass circuit.

DateFeb 2, 2024 (initiated Jan 8, 2024)
StatusOpen, Classified
ReasonNonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.
Root causeProcess control
ActionOn January 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE Removal letters. Customers were instructed to remove all affected RF-32 pumps from use. If product is already in use, the firm advised to not discontinue therapy, but to monitor the patient closely for signs of infection. Customers should quarantine all affected product and return unopened/unexpired affected product to MCP/Getinge. Please contact MCP/Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) to request a
Quantity1090 total
DistributionWorldwide - US Nationwide distribution to the states of CA, IL, MD, MO, NC, NY, OH, OR, PA, TX, UT, VA and the countries of Canada, Germany, Italy, Japan, and Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090515ROTAFLOW CENTRIFUGAL PUMP WITH SOFTLINE COATING, MODEL BO-RF-32 (USA)
Maquet Cardiopulmonary, AG, Dec 18, 2009