Recall Z-0977-2024 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Maquet Medical Systems USA
BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE Coating, Product Code 701047554. For use with an extracorporeal cardiovascular or cardiopulmonary bypass circuit.
| Date | Feb 2, 2024 (initiated Jan 8, 2024) |
| Status | Open, Classified |
| Reason | Nonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis. |
| Root cause | Process control |
| Action | On January 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE Removal letters. Customers were instructed to remove all affected RF-32 pumps from use. If product is already in use, the firm advised to not discontinue therapy, but to monitor the patient closely for signs of infection. Customers should quarantine all affected product and return unopened/unexpired affected product to MCP/Getinge. Please contact MCP/Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) to request a |
| Quantity | 1090 total |
| Distribution | Worldwide - US Nationwide distribution to the states of CA, IL, MD, MO, NC, NY, OH, OR, PA, TX, UT, VA and the countries of Canada, Germany, Italy, Japan, and Netherlands. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K080470 | ROTA FLOW CENTRIFUGAL PUMP WITH BIOLINE COATING, MODELS BE-RF-32, BEQ-RF-32 Maquet Cardiopulmonary, AG, Sep 12, 2008 |