Recall Z-1127-2024 ยท OJH Orthopedic Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline Kits, trays, and packs labeled as follows: a) ANTERIOR PACK, REF DYNJ64814A; b) ARTHROGRAM TRAY, REF DYNDH1629; c) CVL INSERTION KIT-LF, REF DYNJ48587A; d) EXTREMITY CDS, REF CDS983313I; e) EXTREMITY CDS, REF DYNJ902004D; f) EXTREMITY CDS, REF DYNJ902004F; g) EXTREMITY IV, REF CMPJ06040D; h) EXTREMITY IV, REF CMPJ06040F; i) EXTREMITY PROCEDURE, REF CDS981656I; j) FACIAL PLASTY P
| Date | Feb 25, 2024 (initiated Nov 15, 2023) |
| Status | Open, Classified |
| Reason | Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile. |
| Root cause | Nonconforming Material/Component |
| Action | MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different |
| Quantity | 14280 units |
| Distribution | Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.