ClearanceBrief
Recall Z-1076-2024 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Maquet Medical Systems USA

Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.

DateFeb 13, 2024 (initiated Jan 16, 2024)
StatusOpen, Classified
ReasonThe Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
Root causeNonconforming Material/Component
ActionOn January 16, 2024, the recalling firm began notifying affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were instructed to dispose of any Emergency Priming Line from a previously opened, affected HLS set. Customers should only use the Emergency Priming Line from a newly opened HLS Set. Device users should always have an HLS Set ready as a backup during therapy. The remaining components of the backup HLS Set must be disposed of after extraction of the Emergency Priming Line. Please reach out to Getinge Customer Support for the issuance of a replacement HLS Set at no
Quantity16520 worldwide; 6657 US
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112360QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIES
Maquet Cardiopulmonary, AG, Sep 15, 2011