ClearanceBrief
Recall Z-1211-2024 ยท FDF Colonoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

Colonoscope, Model Number PCF-H190L.

DateFeb 29, 2024 (initiated Jan 11, 2024)
StatusOpen, Classified
ReasonSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Root causeNonconforming Material/Component
ActionOn January 11, 2024, the firm notified customers of the issue via URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect your inventory and identify any products of the model and lots subject to this action. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and stop using the device. Customers were given detailed instructions on how to identify affected product. Our Olympus Customer Solutions team will contact you to arrange a no-charge return, and a loaner as necessary. Olympus will perform an inspection of your de
Quantity31
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112680EVIS EXERA III VIDEO SYSTEM
Olympus Medical Systems Corporation, Feb 16, 2012