Recall Z-1078-2024 ยท FZW Expander, Surgical, Skin Graft

Zimmer Surgical Inc

3:1 Dermacarrier, Model Number 00219501300, skin graft carrier

DateFeb 13, 2024 (initiated Jan 2, 2024)
StatusOpen, Classified
ReasonDue to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary.
Root causeUnder Investigation by firm
ActionOn January 2, 2024, the firm notified affected customers of the issue through URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to immediately notify affected personnel of the recall. Customers should also immediately locate and quarantine affected product in inventory. The customer may return the product to Zimmer Biomet, or a Zimmer Biomet sales representative may remove and return the product on the customer's behalf. If product has been further distributed, please notify your customers of the recall and ensure documentation. If you have further questions or concerns after
Quantity76 boxes (1,520 units) US; 521 boxes (10,420 units) OUS
DistributionWorldwide - US Nationwide distribution including in the states of CA, DE, FL, GA, IA, IL, IN, KS, LA, MI, MO, MT, NC, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV and the countries of AUTRALIA, CANADA, DNIPRO OBLAS, HONG KONG, INDIA, JAPAN, MEXICO, CDMX, NETHERLANDS, SEOUL KOREA, SINGAPORE,