Recall Z-1169-2024 ยท LWK Razor, Surgical

3M Company - Health Care Business

3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L

DateFeb 21, 2024 (initiated Jan 12, 2024)
StatusOpen, Classified
Reason3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41C (105.8F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes
Root causeUnder Investigation by firm
Action3M issued a Medical Device Safety Notification - First Notice to its consignees on 01/12/2024 via USPS certified mail and email. The notice explained the issue and its decision to notify its consignees of an added caution statement regarding the use of the product: Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41C (105.8F). " Do not leave the clipper blade applied to the same location on the patie
Quantity52,833,750 blades
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.