Recall Z-1064-2024 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

Vyaire Medical

AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017

DateFeb 25, 2024 (initiated Jan 10, 2024)
StatusOpen, Classified
ReasonRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Root causeProcess control
ActionVyaire issued an URGENT: Customer notification Letter to its consignees on 01/10/2024 via letter delivered by UPS Next Day Air. The notice explained the problem, complaints received, and requested the consignee discard or destroy all affected devices. If the consignee further distributed the product, they were directed to forward a copy of the notice to those parties.
Quantity1,020,330 units
DistributionWorldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.