ClearanceBrief
Recall Z-1146-2024 ยท FXX Mask, Surgical

AMD Medicom Inc.

HENRY SCHEIN CRITERION EARLOOP MASKS LEVEL 3 (BLUE) (item code 570-2448)

DateFeb 16, 2024 (initiated Dec 21, 2023)
StatusOpen, Classified
ReasonLevel 3 masks were produced on alternate non-validated production equipment that were not included as part of the initial qualification of the products.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Recall notification letters dated 12/21/2023 were sent to customers. AMD Medicom requests that you discontinue distribution/use of and return all products from the catalog/lot number referenced above, for a credit. Our records indicate that you have purchased one or more of the referenced products for re-sale to your customers. You should complete the attached Acknowledgment Form for inventory you have received and/or which is still in stock. AMD Medicom requests that you complete and return the Acknowledgment Form enclosed with this letter, even if you do not have the ab
Quantity720 cases (10 boxes/case and 50 masks/box)
DistributionUS Nationwide distribution in the states of FL, IA, IN, PA, NV, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051291NON-STERILE SURGICAL MASK
A.R. Medicom, Inc., Sep 20, 2005