Recall Z-1070-2024 ยท CBK Ventilator, Continuous, Facility Use

Ventec Life Systems, Inc.

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

DateFeb 23, 2024 (initiated Jan 9, 2024)
StatusOpen, Classified
ReasonThe bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
Root causeProcess control
ActionOn 12/22/2023, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION via Direct overnight mail, or FedEx, or email, or hand delivered, to customers informing them that Ventec Life Systems is recalling a specific lot (listed below) of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) since the bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Customer are instructed to: 1. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 2. Stop use of the affected Pa
Quantity250 breathing circuits
DistributionU.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162877VOCSN Unified Respiratory System
Ventec Life Systems, Inc., Apr 7, 2017