ClearanceBrief
Recall Z-0951-2024 ยท LNI System, Nuclear Magnetic Resonance Spectroscopic

Philips North America

Ingenia 3.0T CX. Model (REF) Numbers 781271, 782105.

DateFeb 1, 2024 (initiated Jan 3, 2024)
StatusOpen, Classified
ReasonThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Root causeProcess control
ActionOn January 4, 2024, the firm began notification of affected customers via URGENT Medical Device Correction letters. Customers were provided with precautions to take when using the device, including inspection of the QBC seal for separation. If the QBC seal is found to be loose, or becomes loose during a patient scan, customers should immediately stop use and contact their local Philips service representative. Customers should also ensure that all users of the device are aware of the issue and retain the letter with affected systems until a solution is installed. Philips will contact customer
Quantity1 US; 347 OUS
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei
Philips Medical Systems Nederland B.V., Mar 3, 2022