Recall Z-1054-2024 ยท JAK System, X-Ray, Tomography, Computed

PHILIPS MEDICAL SYSTEMS

Spectral CT on Rails: Software Version 5.1.0.X, model 728334.

DateFeb 8, 2024 (initiated Dec 13, 2023)
StatusOpen, Classified
ReasonA software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Correction notice (2023-PD-CTAMI-017) dated December 13, 2023 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.
QuantityUS: 1
DistributionWorldwide - US Nationwide distribution in the states of AR, AZ, GA, HI, IN, KY, MA, MD, MN, NY, OH, PA, TX & WV; the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Germany, Israel, Italy, Japan, Macao, Malaysia, Netherlands, Norway, Palestine, Panama, Portugal, Singapore,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212875Spectral CT on Rails
Philips Medical Systems Nederland B.V., May 13, 2022