Recall Z-0867-2024 ยท HCF Instrument, Ligature Passing And Knot Tying

Aesculap Inc

MB215R - MILLS DUROGRIP TC Micro Needle Holder, straight, 8:, (203 mm), "TC MILLS NDLE HOLDRRND HNDLSTR203MM"

DateJan 31, 2024 (initiated Dec 29, 2023)
StatusOpen, Classified
ReasonNeedle Holder MB215R was inadvertently distributed as MB362R. Similarly, MB362R was distributed as MB215R during the affected distribution date range. The issue may lead to a short delay in procedure.
Root causeNonconforming Material/Component
ActionOn December 29, 2023, the firm notified affected customers via URGENT MEDICAL DEVICE RECALL NOTIFICATION letters. Customers were instructed to cease use and quarantine all product subject to recall. Once Aesculap Customer Support receives the customer's acknowledgement form, they will contact the customer to arrange for credit and to have the current inventory returned.
Quantity60
DistributionUS Nationwide distribution in the states of CA, CO, FL, ID, KY, MD, MO, MT, NV, NY, OR, PA, RI, TN, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.