Recall Z-1144-2024 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10/10 722005 722011 722027; Allura Xper FD20 722006 722012 722028; Allura Xper FD20 Biplane 722008 722013; Allura Xper FD20 Biplane OR Table 72

DateFeb 16, 2024 (initiated Jan 2, 2024)
StatusOpen, Classified
ReasonFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Root causeProcess design
ActionOn January 2, 2024, the firm notified affected consignees via URGENT Medical Device Correction letters. These letters covered issues with three components of the PCs that may not perform as intended and which may therefore lead to delay in treatment: 1) DIMMs, 2) Disk Bay, and 3) Framegrabber Card. This recall record is for the Framegrabber Card issue. Customers should circulate the notice to all users and establish an emergency protocol for if you experience the issue during a procedure. Customers were provided with monitoring activities to undertake while waiting for their system to be co
Quantity1,993 US; 2,662 OUS
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K130638ALLURACLARITY XPER FD SERIES X-RAY SYSTEM
Philips Medical Systems Nederland B.V., Jun 28, 2013
K130842ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
Philips Medical Systems Nederland B.V., May 21, 2013
K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014
K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014
K162859Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Dec 2, 2016