Recall Z-1152-2024 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: XRay PC, Suite PC, and FlexViewing PC Azurion Model Name - Model Number: Azurion 3 M12 722063 722221; Azurion 3 M15 722064 722222; Azurion 5 M20 722228; Azurion 7 B12 722067 722225; Azurion 7 B20 722068 722226; Azurion 7 M12 722078 722223; Azurion 7 M20

DateFeb 16, 2024 (initiated Jan 2, 2024)
StatusOpen, Classified
ReasonDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Root causeProcess design
ActionOn January 2, 2024, the firm notified affected consignees via" URGENT Medical Device Correction" letters. These letters covered issues with three components of the PCs that may not perform as intended and which may therefore lead to delay in treatment: 1) DIMMs, 2) Disk Bay, and 3) Framegrabber Card. This recall record is for the Disk Bay issue. Customers should circulate the notice to all users and establish an emergency protocol for if you experience the issue during a procedure. Customers were provided with monitoring activities to undertake while waiting for their system to be correcte
Quantity13, 133 (3,408 US; 9,725 ROW)
DistributionWorldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and OUS (International) to countries of: Af
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020