ClearanceBrief
Recall Z-1072-2024 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America

ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref: (1)706100 (2)706110

DateFeb 9, 2024 (initiated Dec 15, 2023)
StatusOpen, Classified
ReasonIssue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and expor
Root causeSoftware design
Action***UPDATE*** 19 ProxiDiagnost N90 Systems (18 US & 1 OUS) added to scope as these systems were produced prior to implementation of the solution in forward production. The newly identified consignees notified on 20 Nov 2024*** Philips issued URGENT Medical Device Correction ProxiDiagnost N90 1.1 and ProxiDiagnost Upgrade on 12/15/2024. Letter states reason for recall, health risk and action to take: For Issues 1 and 2: " To avoid previous patient data from being represented as current patient data, ensure all images have been exported from the previous RAD exam before starting a new examinat
Quantity151 units; 138 units US; 13 units OUS
DistributionNationwide Foreign: Australia Germany India Italy Malaysia Panama Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212837ProxiDiagnost N90
Philips Medical Systems Dmc GmbH, Sep 21, 2021