Recall Z-1053-2024 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Maquet Medical Systems USA

Bubble Sensor (REF: 70105.5720)

DateFeb 8, 2024 (initiated Dec 28, 2023)
StatusOpen, Classified
ReasonThe Bubble Sensor optional accessory to the CARDIOHELP-i extracorporeal life support system includes an incorrect UDI Device Identifier (DI) provided with the label attached to the device.
Root causeLabeling mix-ups
ActionAn Urgent Medical Device Correction Letter was sent VIA Fed-Ex 2-day delivery with signature proof of delivery to all consignees with the following instructions: Actions to be taken by the Customer: - Please examine your inventory immediately to determine if you have any affected product in your inventory. - Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE - CORRECTION RESPONSE FORM to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to BubbleSe
Quantity274 units (70 US, 204 OUS)
DistributionAZ, CA, CO, FL, GA, IA, IL, MN, NE, PA, TX, WA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133598CARDIOHELP SYSTEM
Maquet Cardiopulmonary, AG, May 21, 2014