ClearanceBrief
Recall Z-1542-2024 ยท FRN Pump, Infusion

Smiths Medical ASD Inc.

Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E

DateApr 15, 2024 (initiated Dec 19, 2023)
StatusOpen, Classified
ReasonMedfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Root causeProcess control
ActionSmiths Medical issued an Updated UREGNT MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2023 via US mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: "Customer Required Actions - Medfusion Infusion Pumps: 1. Locate any affected Medfusion Syringe Pumps that may be in your possession by referring to the list of affected devices included with the Response Form. This list includes any specific pump model/serial number(s) your organization purchased that were manufactured or serviced with potentially affected Barrel Clamp Gu
Quantity27 pumps
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040899MEDEX 3000 SERIES MR SYRINGE INFUSION PUMP
Medex, Inc., Apr 30, 2004